Strona 1 - Programmer Manual

xxMSO2000B, DPO2000B, MSO2000 and DPO2000Series OscilloscopesZZZProgrammer Manual*P077073800*077-0738-00

Strona 2

Getting StartedTable 1-2: USB Device ParametersParameter ValueManufacturerID0x0699 (decimal 1689)Product ID0x399 DPO2002B (decimal 921)0x39A MSO2002B

Strona 3

Commands Listed in Alphabetical OrderConditionsThis command requires a DPO2COMP application module.GroupBusSyntaxBUS:B<x>:RS232C:DISplaymo de {F

Strona 4 - Tektronix

Commands Listed in Alphabetical OrderSyntaxBUS:B<x>:RS232C:POLarity {NORMal|INVERTed}BUS:B<x>:RS232C:POLarity?ArgumentsNORMal sets the RS2

Strona 5 - Table of Contents

Commands Listed in Alphabetical OrderBUS:B<x>:SPI:BITOrderSets or returns the bit order for the specified SPI bus.ConditionsThis command requires

Strona 6

Commands Listed in Alphabetical OrderGroupBusSyntaxBUS:B<x>:SPI{:CLOCK|:SCLK} :SOUrce {CH1|CH2|CH3|CH4|D0|D1|D2|D3|D4|D5|D6|D7|D8 |D9|D10|D11|D1

Strona 7

Commands Listed in Alphabetical OrderArgumentsCH1-CH4 is the channel to use as the SPI MISO source.D0-D15 is the digital channel to use as the SPI MIS

Strona 8 - Connecting via Ethernet

Commands Listed in Alphabetical OrderConditionsThis command requires a DPO2EMBD application module.GroupBusSyntaxBUS:B<x>:SPI:DATA:SIZe <NR1&

Strona 9 - Connecting via USB

Commands Listed in Alphabetical OrderGroupBusSyntaxBUS:B<x>:SPI:IDLETime <NR3>BUS:B<x>:SPI:IDLETime?Arguments<NR3> specifies th

Strona 10 - Connecting via GPIB

Commands Listed in Alphabetical OrderArgumentsCH1–CH4 is the channel to use as the SPI SS source.D0–D15 is the digital channel to use as the SPI SS so

Strona 11

Commands Listed in Alphabetical OrderBUS:LOWerthreshold:CH<x>Sets the lower threshold for each channel. This applies to all searchand trigger ty

Strona 12 - Using TekVISA

Commands Listed in Alphabetical OrderArgumentsECL specifies a presetECL high level of –1.3 V.TTL specifies a TTL preset high level of 1.4 V.<NR3>

Strona 13 - Software

Getting Started1. Use the optional5VDCpower adapter connected to the 5 VDCpower inputon the Adapter.2. Use an appropriate USB cable connected to a pow

Strona 14

Commands Listed in Alphabetical OrderArguments<NR3> specifiesthe threshold in volts.ECL specifies a preset ECL high level of –1.3 V.TTL specifies a

Strona 15 - Documentation

Commands Listed in Alphabetical Order*CAL? (Query Only)Performs an internal self-calibration and returns the oscilloscope calibration status.NOTE. Dis

Strona 16 - Getting Started

Commands Listed in Alphabetical OrderCALibrate:FACtory:STATus? (Query Only)Returns the factory calibration status value saved in nonvolatile memory.Gr

Strona 17 - Command Syntax

Commands Listed in Alphabetical OrderCALibrate:INTERNal:STATus? (Query Only)Returns the current status of the oscilloscope internal signal path calibr

Strona 18 - Commands

Commands Listed in Alphabetical OrderCALibrate:RESults:FACtory? (Query Only)Returns the status of internal and factory calibration, without performing

Strona 19 - [:]<Header>

Commands Listed in Alphabetical OrderCH<x>? (QueryOnly)Returns the vertical parameters for channel <x>, where x is the channel number.Grou

Strona 20 - Command E ntry

Commands Listed in Alphabetical OrderArgumentsAC sets channel<x> to AC coupling.DC sets channel <x> to DC coupling.GND sets channel<x&g

Strona 21 - Terminating

Commands Listed in Alphabetical OrderSyntaxCH<x>:IMPedanceCH<x>:IMPedance?CH<x>:INVertSets or returns the invert function for channe

Strona 22 - Constructed Mnemonics

Commands Listed in Alphabetical OrderCH<x>:OFFSetSets or returns the vertical offset for channel <x>, where x is the channel number.This c

Strona 23 - Argument Types

Commands Listed in Alphabetical Orderif the positionfor Channel 3 is set to 2.0 and the offset is set to 3.0, then inputsignals equal to 3.0 units are

Strona 24

Getting StartedNOTE. The CD that your oscilloscope shipped with contains additional tools forefficient connectivity between your oscilloscope and your

Strona 25

Commands Listed in Alphabetical OrderArgumentsExecute auto zeros the probe.CH<x>:PRObe:COMMAND (No Query Form)Sets the state of the p robe contr

Strona 26

Commands Listed in Alphabetical OrderNOTE. This command will returnPASSEDfor probes that do not support degaussoperations.GroupVerticalSyntaxCH<x&g

Strona 27 - Command Groups

Commands Listed in Alphabetical OrderGroupVerticalSyntaxCH<x>:PRObe:GAIN <NR3>CH<x>:PRObe:GAIN?Related CommandsCH<x>:SCAleArgu

Strona 28 - Alias Command

Commands Listed in Alphabetical OrderExamplesCH1:PROBE:ID:SERNUMBER?might return :CH1:PROBE:ID:SERNUMBER"B010289"indicating that the serial

Strona 29 - Bus Command Gr

Commands Listed in Alphabetical OrderGroupVerticalSyntaxCH<x>:PRObe:SIGnal {BYPass|PASS}CH<x>:PRObe:SIGnal?ArgumentsBYPass sets the probe

Strona 30 - Table2-15:BusCommands(cont.)

Commands Listed in Alphabetical OrderSyntaxCH<x>:SCAle <NR3>CH<x>:SCAle?Related CommandsCH<x>:OFFSet, CH<x>:POSition, RE

Strona 31

Commands Listed in Alphabetical OrderCH<x>:YUNitsSets or returns the units for the channel specified by <x>, where x is the channelnumber.

Strona 32 - Calibration a

Commands Listed in Alphabetical Order*CLS can s uppress a Service Request that is to be generated by an * OPC. Thiswill happen if a single sequence ac

Strona 33 - Cursor Comman

Commands Listed in Alphabetical OrderArgumentsOFF removes thecursors from the display but does not change the cursor type.SCREEN specifies both horizon

Strona 34 - Display Command Group

Commands Listed in Alphabetical OrderRelated CommandsCURSor:HBArs:UNItsReturnsA floating point value with an exponent.ExamplesCURSOR:HBARS:DELTA? might

Strona 35 - Ethernet Command Group

Getting Started1. Install the Tektronix OpenChoice Desktop software package, either from theCD that came with your instrument or from the Tektronix we

Strona 36 - File System Command Group

Commands Listed in Alphabetical OrderArgumentsBASe selects the vertical units for the selected waveform.PERcent selects ratio cursors.ExamplesCURSOR:H

Strona 37 - FilterVu Command Group

Commands Listed in Alphabetical OrderSyntaxCURSor:MODe {TRACk|INDependent}CURSor:MODe?ArgumentsTRACk ties the navigational functionality of the two cu

Strona 38 - Hard Copy Command Group

Commands Listed in Alphabetical OrderCURSor:VBArs:DELTa? (Query Only)Returns the horizontal difference between the two v ertical bar cursors. The unit

Strona 39 - Mark Command Group

Commands Listed in Alphabetical OrderCURSor:VBArs:POSITIO N<x>Sets or returns the horizontal position for the specified vertical bar cursor. Thec

Strona 40 - Table 2-24: Mar

Commands Listed in Alphabetical OrderReturnsSECONDS, HERTZ, DEGREES,orPERCENT, depending on the current vertical barcursor units.ExamplesCURSOR:VBARS:

Strona 41 - Math Command G

Commands Listed in Alphabetical OrderSyntaxCURSor:VBArs:VDELTa?Related CommandsCURSor:HBArs:UNItsReturns<NR3> indicates the horizontal differenc

Strona 42 - Measurement Command Group

Commands Listed in Alphabetical OrderCURSor:XY:POLar:THETA:DELta? (Query Only)Returns the XY cursor polar angle delta.GroupCursorSyntaxCURSor:XY:POLar

Strona 43

Commands Listed in Alphabetical OrderCURSor:XY:PRODUCT:POSITION<x>? (Query Only)Returns the position of the X or Y cursor used to calculate the

Strona 44

Commands Listed in Alphabetical OrderCURSor:XY:RATIO:UNIts? (Query Only)Returns the cursor X and cursor Y units for the ratio measurement.GroupCursorS

Strona 45 - Miscellaneous Command Gr

Commands Listed in Alphabetical OrderCURSor:XY:RECTangular:Y:DELta? (Query Only)Returns The cursor Y delta value in rectangular coordinates.GroupCurso

Strona 46 - PictBridge Command Group

Getting StartedTo set up e*Scope communications between your oscilloscope and a n etworkedcomputer:1. With the DPO2CONN Connectivity Module installed,

Strona 47

Commands Listed in Alphabetical Orderpreview parameters. The user should always follow acquisition parameter changeswith a single sequence OPC command

Strona 48

Commands Listed in Alphabetical Order<data> is the curve data.<newline> is a single byte new line character at the end of the data.<asc

Strona 49 - Search Comman

Commands Listed in Alphabetical OrderExamplesDATA? might return:DATA:DESTINATION REF1:ENCDG RIBINARY;SOURCECH1;START 1;STOP 500;WIDTH 1DATA INIT initi

Strona 50 - Search Commands (cont.)

Commands Listed in Alphabetical OrderSyntaxDATa:COMPosition:AVAILable?ReturnsCOMPOSITE_YT for sample data.COMPOSITE_ENV for envelope or peak-detect da

Strona 51

Commands Listed in Alphabetical OrderGroupWaveform TransferSyntaxDATa:ENCdg{ASCIi|FAStest|RIBinary|R PBinary|SRIbinary|SRPbinary}DATa:ENCdg?Related Co

Strona 52

Commands Listed in Alphabetical OrderExamplesDATA:ENCDG? might return:DATa:ENCDG SRPBINARY for the format of theoutgoing w aveform data .DATA:ENCDG RP

Strona 53

Commands Listed in Alphabetical OrderGroupWaveform TransferSyntaxDATa:SOUrce {CH1|CH2|CH3|CH4|MATH|REF1 |REF2|D0|D1|D2|D3|D4|D5|D6|D7|D 8|D9|D10|D11|D

Strona 54 - Trigger Command G roup

Commands Listed in Alphabetical OrderExamplesDATA:SOURCE? might return :DATA:SOURC E REF2 indicating that the source forthe waveform d ata which is tr

Strona 55

Commands Listed in Alphabetical OrderGroupWaveform TransferSyntaxDATa:STARt <NR1>DATa:STARt?Related CommandsCURVe, DATa, DATa:STOP, WFMInpre:NR_

Strona 56 - Serial bus application

Commands Listed in Alphabetical OrderArguments<NR1> is the last data point that will be transferred, which ranges from 1 tothe record length. If

Strona 57 - Table 2-31: Tri

Getting Started4. On the menu bar of your PC’s web browser, type in the oscilloscope’sInstrument IP address and press Enter.5. You should now see the

Strona 58

Commands Listed in Alphabetical OrderSyntaxDATE <Q String>DATE?Related CommandsTIMEArguments<QString> is a date in the form "yyyy-mm-

Strona 59

Commands Listed in Alphabetical OrderDESESets or returns the bits in the Device Event Status Enable Register (DESER). TheDESER is the mask that determ

Strona 60

Commands Listed in Alphabetical OrderArgumentsALWAYS continues looping until the self tests (diagnostics) are stopped via thefront panel or by an osci

Strona 61

Commands Listed in Alphabetical OrderDIAg:RESUlt:FLAg? (Query Only)Returns the pass/fail status from the last self-test sequence exec ution. U se this

Strona 62

Commands Listed in Alphabetical OrderDIAg:SELect (No Query Form)Sets the type of diagnostics grouping.GroupCalibration a nd DiagnosticSyntaxDIAg:SELec

Strona 63 - Vertical Command Group

Commands Listed in Alphabetical OrderExamplesDIAG:SELECT:CPUsets the oscilloscope to run just CPU tests.DIAg:STATE (No Query Form)This command starts

Strona 64

Commands Listed in Alphabetical OrderRelated CommandsDATE, TIMEArgumentsDATEOnly enables the display of date.TIMEOnly enables the display of time.DATE

Strona 65

Commands Listed in Alphabetical OrderSyntaxDISplay:FORMat {YT|XY}DISplay:FORMat?ArgumentsYT sets the display to a voltage versus time format and is th

Strona 66

Commands Listed in Alphabetical OrderSyntaxDISplay:GRAticule {CRO SSHair|FRAme|FUL l|GRId}DISplay:GRAticule?ArgumentsCROSSHair specifies a frame and cr

Strona 67

Commands Listed in Alphabetical OrderDISplay:INTENSITy:GLITchSets the intensity of the glitch capture background of the waveform display. Theintensity

Strona 68 - Waveform Data and Record

Getting Started1-10 MSO2000B, DPO2000B, MSO2000 and DPO2000 Series Oscilloscopes Programmer Manual

Strona 69

Commands Listed in Alphabetical OrderGroupDisplaySyntaxDISplay:INTENSITy:WAVEfor m <NR1>DISplay:INTENSITy:WAVEfor m?Arguments<NR1> is the

Strona 70 - Data to the Oscilloscope

Commands Listed in Alphabetical OrderMINImum specifies that the waveform persistence is set to the minimum valueof 0.0E0.ExamplesDISPLAY:PERSISTENCE 3s

Strona 71

Commands Listed in Alphabetical OrderGroupMSOSyntaxD<x>:POSition <NR3>D<x>:POSition?Arguments<NR3> is the position of the digi

Strona 72

Commands Listed in Alphabetical OrderRelated Commands*CLS, DESE, *ESR?, EVENT?, EVMsg?, *SRE, *STB?Arguments<NR1> specifies the binary bits of th

Strona 73 - Zoom C om mand Group

Commands Listed in Alphabetical OrderSyntaxETHERnet:DHCPbootp {ON |OFF}ETHERnet:DHCPbootp?ArgumentsON enables the oscilloscope t o search the network

Strona 74

Commands Listed in Alphabetical OrderArguments<QString> is the network domain name, enclosed in quotes.ExamplesETHERNET:DOMAINNAME “Alpha1.Mycor

Strona 75 - ACQuire:MODe

Commands Listed in Alphabetical OrderSyntaxETHERnet:HTTPPort <QSt ring>ETHERnet:HTTPPort?Arguments<QString> is an integer port number, enc

Strona 76 - ACQuire:NUMACq? (Query O nly)

Commands Listed in Alphabetical OrderExamplesETHERNET:NAME“labscope1” sets the oscilloscope's network name.ETHERnet:PASSWordSets or returns the H

Strona 77 - ACQuire:STATE

Commands Listed in Alphabetical OrderReturnsOK is returned if the computer at the gateway IP address answers.NORESPOnsE is returned if the computer at

Strona 78

Commands Listed in Alphabetical OrderEVMsg? (QueryOnly)Removes a single event code from the Event Queue that is associated with theresults of the last

Strona 79 - ACQuire:STOP

Command SyntaxYou can control the operations and functions of the oscilloscope through theEthernet port or the USB 2.0 device port using commands and

Strona 80 - ALIas:DEFine

Commands Listed in Alphabetical OrderFACtory (No Query Form)Resets the oscilloscope to its factory default settings.This command does the following:Cl

Strona 81 - ALIas:DELEte (No Query Form)

Commands Listed in Alphabetical OrderFILESystemReturns the directory listing of the current working directory and the number ofbytes of free space ava

Strona 82 - ALIas[:STATE]

Commands Listed in Alphabetical OrderSyntaxFILESystem:CWD {<new working directory p ath>}Arguments<new wo rking directory path> is a quote

Strona 83 - AUTOSet (No Query Form)

Commands Listed in Alphabetical OrderGroupFile SystemSyntaxFILESystem:DIR?Related CommandsFILESystem:CWD, FILESystem:MKDirArgumentsNoneReturnsFILESyst

Strona 84 - AUXin:PRObe

Commands Listed in Alphabetical OrderGroupFile SystemSyntaxFILESystem:FREESpace?Related CommandsFILESystem:FREESpace?, FILESystem:CWDFILESystem:MKDir

Strona 85

Commands Listed in Alphabetical OrderRelated CommandsFILESystem:CWDArguments<QString> is a quoted string that defines the file name and path. If t

Strona 86 - AUXin:PRObe:FORCEDRange

Commands Listed in Alphabetical OrderRelated CommandsFILESystem:CWDArguments<directory path> is a quoted s tr ing that defines the directory name

Strona 87 - AUXin:PRObe:GAIN

Commands Listed in Alphabetical OrderArguments<NR3> is the integer format for the current FilterVu frequency setting.ExamplesFILTERVU:FREQUENCY?

Strona 88 - AUXin:PRObe:SIGnal

Commands Listed in Alphabetical OrderTable 2-37: FPAnel:PRESS argumentsArgument DescriptionACQuireAcquire buttonSAVERecall Save/RecallMenu buttonMEASu

Strona 89 - BUS:B<x>:CAN:PRObe

Commands Listed in Alphabetical OrderTable 2-37: FPAnel:PRESS arguments (cont.)Argument DescriptionBMENU5Screen bottom menu buttonBMENU6Screen bottom

Strona 90 - BUS:B<x>:CAN:SOUrce

Command SyntaxTable 2-2: Command Message ElementsSymbol Meaning<Header>This is the basic command name. If the header ends with a questionmark, t

Strona 91 - BUS:B<x>:DISplay:FORMAt

Commands Listed in Alphabetical Orderclockwise rotation. The magnitude of <n> specifies the amount of the turn, where<n> = 1 represents tur

Strona 92

Commands Listed in Alphabetical OrderRelated Commands*WAI, *CLSArgumentsSTART sends a block of data representing the current screen image to the reque

Strona 93 - BUS:B<x>:LIN:IDFORmat

Commands Listed in Alphabetical OrderArgumentsON or <NR1> ≠ 0turns preview mode on.OFF or <NR1> = 0 turns preview mode off.HEADerSets or r

Strona 94 - BUS:B<x>:LIN:POLARity

Commands Listed in Alphabetical OrderSyntaxHORizontal?ExamplesHORIZONTAL? might return the following horizontalsettings:HORIZONTAL:DIGITAL:RECOR DLENG

Strona 95 - BUS:B<x>:LIN:STANDard

Commands Listed in Alphabetical Order<NR1> = 0 sets the Horizontal Delay Mode to off; any other value sets this modeto on.ExamplesHORIZONTAL:DEL

Strona 96

Commands Listed in Alphabetical OrderSyntaxHORizontal:DIGital:SAMPLERate:MAIN?HORizontal:POSitionSets or returns the horizontal position. If Horizonta

Strona 97 - BUS:B<x>:PARallel:WIDth

Commands Listed in Alphabetical OrderGroupHorizontalSyntaxHORizontal:RECOrdlength <NR1>HORizontal:RECOrdlength?Arguments<NR1> represents t

Strona 98 - BUS:B<x>:RS232C:BITRate

Commands Listed in Alphabetical OrderGroupHorizontalSyntaxHORizontal:SCAle <NR3>HORizontal:SCAle?Arguments<NR3> specifies the range from 2

Strona 99 - 32C:DATABits

Commands Listed in Alphabetical Order*IDN? (Query Only)Returns the oscilloscope identification code.GroupMiscellaneousSyntax*IDN?Related CommandsID?Exa

Strona 100 - BUS:B<x>:RS232C:PARity

Commands Listed in Alphabetical OrderGroupMiscellaneousSyntaxLOCk {A LL|NONe}LOCk?Related CommandsUNLockArgumentsALL disables all front-panel controls

Strona 101

Command SyntaxQueriesQueries cause the oscilloscope to return status or setting information. Querieshave the structure:[:]<Header>[:]<Header&

Strona 102 - :BITOrder

Commands Listed in Alphabetical OrderINTERNAL;:SAVE:ASSIGN:TYPE SETUP;:TRIGGER:A:MODEAUTO;TYPE EDGE;LE V EL 20.0000E-3;LEVEL:CH120.0000E-3;CH2 0.0000;

Strona 103

Commands Listed in Alphabetical OrderSyntaxMARK:CREATE {CH<x>|MATH|B<x>|REF<x>|COLUM N|DIGItal}ArgumentsCH<x> creates the mark

Strona 104 - BUS:B<x>:SPI:DATA:SIZe

Commands Listed in Alphabetical OrderMARK:SELected:END? (Query Only)Returns the end of the selected mark, 0 to 100% of the waveform.GroupMarkSyntaxMAR

Strona 105 - BUS:B<x>:SPI:IDLETime

Commands Listed in Alphabetical OrderMARK:SELected:SOURCE? (Query Only)Returns the source waveform for the selected mark.GroupMarkSyntaxMARK:SELected:

Strona 106

Commands Listed in Alphabetical OrderMARK:TOTal? (Query Only)Returns how many marks are currently in use.GroupMarkSyntaxMARK:TOTal?MATH[1]:LABelSets o

Strona 107 - BUS:B<x>:TYPE

Commands Listed in Alphabetical Order"V";:MATH:HORIZONTAL:SCALE 4.0000E-6;POSITION 50.0000;UNITS"s";:MATH:SPECTRAL:MAG DB;WINDOW H

Strona 108 - BUS:THReshold:CH<x>

Commands Listed in Alphabetical OrderExamplesMATH1:DEFINE "CH1+CH2"adds the Ch 1 waveform and Ch 2 waveform,storing the results in Math 1.MA

Strona 109 - BUS:THReshold:D<x>

Commands Listed in Alphabetical OrderMATH[1]:HORizontal:UNItsReturns the math waveform horizontal measurement unit value.GroupMathSyntaxMATH[1]:HORizo

Strona 110 - BUSY? (Query Only)

Commands Listed in Alphabetical OrderReturnsA value between0 and 1, which represents the fractional portion of the sourcerecord.MATH[1]:SPECTral:GATin

Strona 111 - *CAL? (Query O

Commands Listed in Alphabetical OrderSyntaxMATH[1]:SPECTral:NYQUISTFreq?MATH[1]:SPECTral:WINdowSets or returns the window function for the spectral an

Strona 113 - TERNal:STATus? (Query Only)

Command SyntaxClearing the oscilloscopeYou can clear the Output Queue and reset the oscilloscope to accept a newcommand or query by using the selected

Strona 114 - Sults:FACtory? (Query Only)

Commands Listed in Alphabetical OrderRelated CommandsMATH[1]:DEFineArgumentsDUAL sets the math waveform mode to dual waveform math.FFT sets the math w

Strona 115 - CH<x>:COUPling

Commands Listed in Alphabetical OrderRelated CommandsCH<x>:SCAle, REF<x>:VERTical:SCAleArguments<NR3> is the scale-per-division in t

Strona 116 - CH<x>:IMPedance

Commands Listed in Alphabetical OrderDELAY:DIRECTION FORWAR DS;EDGE1 RISE;EDGE2RISE;:MEASUREMENT:MEAS3:S TATE 1;TYPE PK2PK;UNITS "V";SOURCE1

Strona 117 - CH<x>:LABel

Commands Listed in Alphabetical OrderCURSor limits measurements to the portion of the waveform between the verticalbar cursors, even if they are off s

Strona 118 - CH<x>:POSition

Commands Listed in Alphabetical OrderMEASUrement:IMMed:DELay:DIRectionSets or returns the starting point and direction that determines the delay "

Strona 119

Commands Listed in Alphabetical OrderRelated CommandsMEASUrement:IMMed:SOUrce1MEASUrement:IMMed:SOUrce2Arguments<x> specifies which waveform to u

Strona 120

Commands Listed in Alphabetical OrderExamplesMEASUREMENT:IMMED:SOURCE1 MATH1specifies Math1 as t he immediate measurement source.MEASUREMENT:IMMED:SOUR

Strona 121 - H<x>:PRObe:GAIN

Commands Listed in Alphabetical OrderFor SOUrce2: Sets or returns the waveform to measure "to" when taking a delaymeasurement or phase measu

Strona 122

Commands Listed in Alphabetical OrderCMEan (cycle mean) measures the a rithmetic mean over the first cycle in thewaveform or the first cycle in the gate

Strona 123 - CH<x>:PRObe:SIGnal

Commands Listed in Alphabetical OrderNWIdth (negative width) measurement is the distance (time) between the middlereference (default = 50%) amplitude

Strona 124 - CH<x>:SCAle

Command SyntaxWhen concatenating commands and queries, you must follow these rules:1. Separate completely different headers by a semicolon and by the

Strona 125 - CH<x>:VOLts

Commands Listed in Alphabetical OrderVOLTS, VOLTS SQUARED, SEC, HERTZ, PERCENT, DIVS, SAMPLES,OHMS, AMPS, WATTS, MINUTES, DEGREES, UNKNOWN, AMPSSQUARE

Strona 126 - *CLS (No Query Form)

Commands Listed in Alphabetical OrderMEASUrement:INDICators? (Query Only)Returns all measurement indicator parameters.GroupMeasurementSyntaxMEASUremen

Strona 127 - CURSor:FUNCtion

Commands Listed in Alphabetical Order2 horizontal lines drawn on the graticule. The indicators show where themeasurement specified by MEASUrement:INDIC

Strona 128 - CURSor:HBArs? (Query Only)

Commands Listed in Alphabetical OrderMEASUrement:INDICators:VERT<x>? (Query Only)Returns the value of the specified vertical measurement indicato

Strona 129 - CURSor:HBArs:UNIts

Commands Listed in Alphabetical OrderMEASUrement:MEAS<x>:DELay? (Query Only)Returns the delay measurement parameters for the measurement specifie

Strona 130 - CURSor:MODe

Commands Listed in Alphabetical OrderMEASUREMENT:MEAS3:DELAY:DIRECTION?might return:MEASUREMENT:MEAS3:DELAY:D IRECTION BACKWARDS indicatingthat the cu

Strona 131 - CURSor:VBArs? (Query Only)

Commands Listed in Alphabetical OrderMEASUrement:MEAS<x>:MEAN? (Query Only)Returns the mean value accumulated for this measurement since the las

Strona 132 - :DELTa? (Query Only)

Commands Listed in Alphabetical OrderExamplesMEASUREMENT:MEAS2:SOURCE1 MATH1specifies Math 1 as the measurement2 source.MEASUREMENT:MEAS1:SOURCE1?might

Strona 133 - CURSor:VBArs:UNIts

Commands Listed in Alphabetical OrderFor SOUrce2: Sets or returns the waveform to measure "to" when taking a delaymeasurement or phase measu

Strona 134

Commands Listed in Alphabetical OrderMEASUREMENT:MEAS1:STATE?might return : MEASUREMENT:MEA S1:STATE 0indicating that measurement defined for measureme

Strona 135

Command SyntaxTable 2-4: End of Message TerminatorSymbol Meaning<EOM>Message terminatorThe end-of-message terminator must be the END message (EO

Strona 136 - CURSor:XY:PRODUCT:DELta?

Commands Listed in Alphabetical OrderCARea (cycle area) measures the voltage over time. In other words, it measures, involt-seconds, the area over the

Strona 137 - CURSor:XY:PR

Commands Listed in Alphabetical OrderNOVershoot (negative overshoot) finds the negative overshoot value over theentire waveform or gated region. This m

Strona 138 - TIO:UNIts? (Query Only)

Commands Listed in Alphabetical OrderMEASUREMENT:MEAS1:TYPE?might return :MEASUREMENT:MEAS 1:TYPE RMSindicating that measurement 1 is defined to measur

Strona 139 - CURSor:XY:RE

Commands Listed in Alphabetical OrderMEASUrement:METHodSets or returns the method used to calculate the 0% and 1 00% reference level.GroupMeasurementS

Strona 140 - Arguments

Commands Listed in Alphabetical OrderMEASUrement:REFLevel:ABSolute:HIGHSets or returns the high reference level, and is the upper reference level when

Strona 141 - Examples

Commands Listed in Alphabetical OrderRelated CommandsMEASUrement:REFLevel:METHod, MEASUrement:IMMed:TYPe,MEASUrement:MEAS<x>:TYPeArguments<NR

Strona 142 - DATa:COMPosition

Commands Listed in Alphabetical OrderMEASUrement:REFLevel:ABSolute:MID2Sets or returns the mid reference level for the "to" waveform whentak

Strona 143 - DATa:ENCdg

Commands Listed in Alphabetical OrderMEASUrement:REFLevel:METHodSpecifies or returns the reference level units used for measurement calculations.NOTE.

Strona 144

Commands Listed in Alphabetical OrderSyntaxMEASUrement:REFLevel:PERCent:HIGH <NR3>MEASUrement:REFLevel:PERC ent:HIGH?Related CommandsMEASUrement

Strona 145 - DATa:SOUrce

Commands Listed in Alphabetical OrderMEASUREMENT:REFLEVEL:PERCENT:LOW?might return:MEASUREMENT:REFLEVEL:PERC ENT:LOW 10 indicating that thepercentage

Strona 146

Command SyntaxMeasurement SpecifierMnemonicsCommands can specify whic h measurement to set or query as a mnemonic in theheader. Up to four automated me

Strona 147 - DATa:STARt

Commands Listed in Alphabetical OrderGroupMeasurementSyntaxMEASUrement:REFLevel:PERC ent:MID2 <NR3>MEASUrement:REFLevel:PERC ent:MID2?Related Co

Strona 148 - DATa:STOP

Commands Listed in Alphabetical OrderMESSageThis command sets o r queries message parameters.GroupMiscellaneousSyntaxMESSageMESSage?ExamplesMESSAGE? m

Strona 149 - DATa:WIDth

Commands Listed in Alphabetical Ordercommand automatically sizes the box to fi t the message. All four values ar ereturned in a query.MESSage:CLEAR (No

Strona 150

Commands Listed in Alphabetical OrderThe message string itself is not altered. The entire message can be returned as aquery response regardless of wha

Strona 151 - DIAg:LOOP:OPTion

Commands Listed in Alphabetical OrderIndex 10 Tek blueIndex 11 Bright blueIndex 12UndefinedIndex 13 BlueIndex 14UndefinedIndex 15 Dark blueBit 4If set,

Strona 152 - DIAg:LOOP:OPTion:NTIMes

Commands Listed in Alphabetical OrderExamplesMESSage:SHOW “Hello World”displays “Hello world” in the upper left corner of the box (you can define the b

Strona 153 - DIAg:RESUlt:LOG? (Query Only)

Commands Listed in Alphabetical OrderArguments<QString> is the new password, which can contain up to 16 characters.ExamplesNEWPASS "myp ass

Strona 154 - No Query Form)

Commands Listed in Alphabetical Order*OPC? might return 1 to indicate that all pending OPC operations are finished.PASSWord(No Query Form)Enables the *

Strona 155 - DISplay:CLOCk

Commands Listed in Alphabetical OrderExamplesPICTBRIDGE:DATEPRINT? might return PICTBRIDGE:DATEPRINT DEFLTindicating that the default setting of the p

Strona 156 - DISplay:FORMat

Commands Listed in Alphabetical OrderPICTBridge:IMAGESizeSets or returns the image print size.GroupPictBridgeSyntaxPICTBridge:IMAGESize {DEFLT|IN2P5BY

Strona 157 - DISplay:GRAticule

Command SyntaxTable 2-11: Quoted String ArgumentSymbol Meaning<QString> Quoted string of AS CII textA quoted string can include any character de

Strona 158 - DISplay:INTENSITy:BACKLight

Commands Listed in Alphabetical OrderPICTBridge:PAPERSizeSets the output print paper size.GroupPictBridgeSyntaxPICTBridge:PAPERSize{DEFLT|L|L2|HAGAKIP

Strona 159 - DISplay:INTENSITy:WAVEform

Commands Listed in Alphabetical OrderB2 for a B2 papersize.B3 for a B3 paper size.B4 for a B4 paper size.B5 for a B5 paper size.B6 for a B6 paper size

Strona 160 - DISplay:PERSistence

Commands Listed in Alphabetical OrderPICTBridge:PRINTQualSets or returns the output print quality.GroupPictBridgeSyntaxPICTBridge:PRINTQual {DEFLT|NRM

Strona 161 - D<x>:POSition

Commands Listed in Alphabetical Ordervalue sets the power-on status clear flag to true, enabling the power-on status clearpreventing any SRQ assertion

Strona 162 - D<x>:THREshold

Commands Listed in Alphabetical OrderArguments<NR1> is a valuein the range from 1 to 10, which specifies a saved setup storagelocation.Examples*R

Strona 163 - ETHERnet:DHCPbootp

Commands Listed in Alphabetical OrderGroupSave and RecallSyntaxRECAll:WAVEform <file path>,REF<x>Related CommandsSAVe:WAVEform, FILESystem

Strona 164 - ETHERnet:DOMAINname

Commands Listed in Alphabetical OrderREF<x>:HORizontal:DELay:TIMeSets or returns the horizontal delay time for reference w aveform <x>, wh

Strona 165 - ETHERnet:HTTPPort

Commands Listed in Alphabetical OrderSyntaxREF<x>:LABel <Qstring>REF<x>:LABel?Arguments<Qstring> is an a lpha-numeric string o

Strona 166 - ETHERnet:NAME

Commands Listed in Alphabetical OrderGroupVerticalSyntaxREF<x>:TIMe?ExamplesREF2:TIME? might return “16:54:05”.REF<x>:VERTical:POSitionSet

Strona 167 - ETHERnet:PING (No Query Form)

Commands Listed in Alphabetical OrderEach waveform has a vertical scale parameter. For a signal with constantamplitude, increasing the Scale causes th

Strona 168 - EVENT? (Query Only)

Command SyntaxMSO2000B, DPO2000B, MSO2000 and DPO2000 Series Oscilloscopes Programmer Manual 2-9

Strona 169 - EVQty? (Query Only)

Commands Listed in Alphabetical OrderExamplesREM "This is a comment"is a comment string that the oscilloscope willignore.*RST (No Query Form

Strona 170 - FACtory (No Qu

Commands Listed in Alphabetical OrderRelated Commands*RCL, RECAll:SETUp, SAVe:SETUpArguments<NR1> specifies a location in which to save the state

Strona 171 - FILESystem:CWD

Commands Listed in Alphabetical OrderSAVe:IMAGe (No Query Form)Saves a capture of the screen image into the specified file. Supported imageformats are P

Strona 172 - FILESystem:DIR? (Query Only)

Commands Listed in Alphabetical OrderSAVe:IMAGe:LAYoutSets or returns the layout to use for saved screen images.GroupSave and RecallSyntaxSAVe:IMAGe:L

Strona 173

Commands Listed in Alphabetical OrderSAVe:WAVEform(NoQueryForm)This command saves a specified waveform or all displayed waveforms (excludingserial bus

Strona 174

Commands Listed in Alphabetical Order:SAVE:WAVEFORM:FILEFORMAT SPREADSHEET; :SAVE:WAVEFORM ALL,"E:/test_folder/test1_all. csv"saves all disp

Strona 175

Commands Listed in Alphabetical OrderSAVe:WAVEform:GATIngSpecifies whether save waveform operations should save the entire waveform(NONe) or a specified

Strona 176 - FILTERVu:FREQuency

Commands Listed in Alphabetical OrderSAVE:WAVEFORM:SPREADSHEET:RESOLUTIONREDUc edsets the reducedresolution waveform to be saved to a CSV file.SEARCH?

Strona 177 - FPAnel:PRESS (No Query Form)

Commands Listed in Alphabetical Order:SEARCH:SEARCH1:TRIGGER:A:BUS:B1:I2C:DATA :DIRECTION NOCARE;:SEARCH:SEARCH1:TRIGGER:A :BUS:B2:I2C:DATA:DIRECTION

Strona 178 - Table 2-37: FPA

Commands Listed in Alphabetical Order:SEARCH:SEARCH1:TRIGGER:A:BUS:B2:LIN:ERRT YPE SYNC;:SEARCH:SEARCH1:TRIGGER:A: BUS:B1:PARALLEL:VALUE"XXXXXXXX

Strona 179 - FPAnel:TURN (No Query Form)

Command Syntax2-10 MSO2000B, DPO2000B, MSO2000 and DPO2000 Series Oscilloscopes Programmer Manual

Strona 180 - HARDCopy (No Query Form)

Commands Listed in Alphabetical OrderTHRESHOLD:CH10.0E+0;CH2 0.0E+0;CH3 0.0E+0;CH4 0.0E+0; MATH0.0E+0;REF1 0.0E+0;REF2 0.0E+0;:SEARCH:SEARCH1:STATE 0S

Strona 181 - HARDCopy:INKSaver

Commands Listed in Alphabetical OrderSyntaxSEARCH:SEARCH<x>:TOTAL?Returns<NR1> is the total number of matches.SEARCH:SEARCH<x>:TRIGg

Strona 182 - HORizontal? (Query Only)

Commands Listed in Alphabetical Order:SEARCH:SEARCH1:TRIGGER:A:BUS:B2:I2C:DATA :VALUE "XXXXXXXX";:SEARCH1:TRIGGER:A:BUS:B1 :I2C:DATA:SIZE 1;

Strona 183 - HORizontal:DELay:MODe

Commands Listed in Alphabetical Order:SEARCH:SEARCH1:TRIGGER:A:BUS:B2:LIN:IDEN TIFIER:VALUE"XXXXXX";:SEARCH:SEARCH1:TRIGGER:A: BUS:B1:LIN:ER

Strona 184 - LERate:MAIN? (Query Only)

Commands Listed in Alphabetical OrderConditionsThis command requires a DPO2AUTO application m odule.GroupSearchSyntaxSEARCH:SEARCH<x>:TRIGger: A

Strona 185 - HORizontal:RECOrdlength

Commands Listed in Alphabetical OrderLESSEQual searches for bus data less than or equal to the value specified bySEARCH:SEARCH<x>:TRIGger:A:BUS:B

Strona 186 - HORizontal:SCAle

Commands Listed in Alphabetical OrderSEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:CAN:FRAMEtypeSets or returns the CAN Frame Type to be used if the

Strona 187 - ID? (Query Only)

Commands Listed in Alphabetical OrderSEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:CAN{:IDentifier|:ADDRess}:VALueSets or returns the binary address

Strona 188 - LANGuage

Commands Listed in Alphabetical OrderGroupSearchSyntaxSEARCH:SEARCH<x>:TRIGger: A:BUS:B<x>:I2C:ADDRess:TYPe{GENeralcall|STARtbyte|HS mode|

Strona 189 - *LRN? (Query Only)

Commands Listed in Alphabetical OrderSyntaxSEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:I2C:C ONDition{STARt|STOP|REPEATstart|AC KMISS|ADDRess|DATA

Strona 190 - MARK:CREATE (No Query Form)

Command GroupsThis manual lists the MSO/DPO2000B and MSO/DPO2000 series IEEE488.2commands in two ways. First, it presents them by functional groups. T

Strona 191 - MARK:FREE? (Query Only)

Commands Listed in Alphabetical OrderGroupSearchSyntaxSEARCH:SEARCH<x>:TRIGger: A:BUS:B<x>:I2C:DATa:SIZe <NR1>SEARCH:SEARCH<x>

Strona 192 - MARK:SELecte

Commands Listed in Alphabetical OrderIDANDDATA specifies to search on the identifier and the data.WAKEup specifies to search on wake up.SLEEP specifies to

Strona 193

Commands Listed in Alphabetical OrderSyntaxSEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:LIN:D ATa:QUALifier{LESSThan|MOREThan|EQUal| UNEQual|LESSEQ

Strona 194 - MATH[1]? (Query Only)

Commands Listed in Alphabetical OrderSEARCH:SEARCH1:TRIGGER:A:BUS:B1:LIN:DATA: SIZE?might returnSEARCH:SEARCH1:TRIGGER:A:B US:B1:LIN:DATA:SIZE 1 indic

Strona 195 - MATH[1]:DEFine

Commands Listed in Alphabetical OrderExamplesSEARCH:SEARCH1:TRIGGER:A:BUS:B1:LIN:ERRTY PE?might returnSEARCH:SEARCH1:TRIGGER:A: BUS:B1:LIN:ERRTYPE SYN

Strona 196 - MATH[1]:HORizontal:SCAle

Commands Listed in Alphabetical OrderSEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:RS232C:CONDitionSets or returns the condition for a RS232 trigger

Strona 197 - MATH[1]:HORi

Commands Listed in Alphabetical OrderSEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:RS232C:RX:DATa:VALueSets or returns the binary data string for a

Strona 198 - MATH[1]:SPECTral:MAG

Commands Listed in Alphabetical OrderSyntaxSEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:RS232 C:TX:DATa:VALueSEARCH:SEARCH<x>:TRIGger:A :BUS:

Strona 199 - MATH[1]:TYPe

Commands Listed in Alphabetical OrderSEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:SPI:DATa{:MOSI|:O UT}:VALueSets or returns the binary data string

Strona 200 - MATH[1]:VERTical:SCAle

Commands Listed in Alphabetical OrderGroupSearchSyntaxSEARCH:SEARCH<x>:TRIGger:A :BUS:SOUrce {B1|B2}SEARCH:SEARCH<x>:TRIGger:A :BUS:SOUrce

Strona 201 - MEASUrement? (Query Only)

Command GroupsAlias CommandGroupUse the Alias commands to define new commands as a sequence of standardcommands. You may find this useful when repeatedl

Strona 202 - MEASUrement:GATing

Commands Listed in Alphabetical OrderSEARCH:SEARCH<x>:TRIGger:A:LEVelSets or returns the level for an edge trigger search to deter mine where to

Strona 203

Commands Listed in Alphabetical OrderSyntaxSEARCH:SEARCH<x>:TRIGger:A:LEVel:MATH {TTL|ECL}SEARCH:SEARCH<x>:TRIGger:A :LEVel:MATH?Arguments

Strona 204 - IMMed:DELay:DIRection

Commands Listed in Alphabetical OrderGroupSearchSyntaxSEARCH:SEARCH<x>:TRIGger: A:LOGIc:INPut:CH<x> {HIGH |LOW|X}SEARCH:SEARCH<x>:TR

Strona 205 - MEASUrement:IMMed:SOUrce1

Commands Listed in Alphabetical OrderArgumentsCH<x> specifiesa channel input as the clock source, where <x> = 1, 2, 3, or 4.MATH specifies t

Strona 206 - MEASUrement:IMMed:SOUrce2

Commands Listed in Alphabetical OrderSEARCH:SEARCH<x>:TRIGger:A:LOGIc:INPut:REF<x>Sets or returns the Boolean logic criteria for a logic t

Strona 207 - MEASUrement:IMMed:TYPe

Commands Listed in Alphabetical OrderSyntaxSEARCH:SEARCH<x>:TRIGger:A:LOGIc:PATtern: INPut:D<x>{HIGH|LOW|X}SEARCH:SEARCH<x>:TRIGger:

Strona 208

Commands Listed in Alphabetical OrderSEARCH:SEARCH<x>:TRIGger:A:LOGIc:PATtern:WHEnSets or returns the condition for generating a logic pattern t

Strona 209

Commands Listed in Alphabetical OrderSEARCH:SEARCH<x>:TRIGger:A:LOGIc:PATtern:WHEn:MORELimitSets or returns the minimum time that the selected p

Strona 210

Commands Listed in Alphabetical OrderArgumentsTTL specifies a presetTTL high level of 1.4 V.ECL specifies a preset ECL high level of –1.3 V.<NR3>

Strona 211 - INDICators? (Query Only)

Commands Listed in Alphabetical OrderSEARCH:SEARCH<x>:TRIGger:A:LOWerthreshold:MATHSets or returns the math waveform lower threshold to determin

Strona 212 - MEASUrement:INDICators:STATE

Command GroupsBus Command GroupUse the Bus commands when working with serial bus measurements.Install the DPO2EMBD application module when working wit

Strona 213

Commands Listed in Alphabetical OrderGroupSearchSyntaxSEARCH:SEARCH<x>:TRIGger: A:PULSEWidth:POLarity{NEGative|POSitive}SEARCH:SEARCH<x>:T

Strona 214

Commands Listed in Alphabetical OrderMOREthan places a mark if the pulse width is more than the time set by theSEARCH:SEARCH<x>:TRIGger:A:PULSEW

Strona 215

Commands Listed in Alphabetical OrderSEARCH:SEARCH<x>:TRIGger:A:RUNT:SOUrceSets or returns the source setting for a runt trigger search to deter

Strona 216

Commands Listed in Alphabetical OrderSEARCH:SEARCH<x>:TRIGger:A:RUNT:WIDthSets or returns the width setting for a runt trigger search to determi

Strona 217 - MEASUremen

Commands Listed in Alphabetical OrderArgumentsCH1–CH4 specifies an input channel as the edge source.MATH specifies the math waveform as the search sourc

Strona 218

Commands Listed in Alphabetical OrderMATH specifies the math waveform as the search s ource.REF specifies the reference waveform as the search source.D0

Strona 219

Commands Listed in Alphabetical OrderGroupSearchSyntaxSEARCH:SEARCH<x>:TRIGger: A:SETHold:SETTime <NR3>SEARCH:SEARCH<x>:TRIGger: A:S

Strona 220

Commands Listed in Alphabetical OrderSEARCH:SEARCH<x>:TRIGger:A:SETHold:THReshold:REF<x>Sets or returns the trigger search setup and hold

Strona 221

Commands Listed in Alphabetical OrderArgumentsPOSitive specifies that a pulse edge must traverse from the lower (mostnegative) to higher (most positive

Strona 222

Commands Listed in Alphabetical OrderUNEQual sets the trigger to occur when the transitioning signal is not equal tothe set volts/second rate ±5%.SEAR

Strona 223

MSO2000B, DPO2000B, MSO2000 and DPO2000Series OscilloscopesZZZProgrammer ManualxxRevision Awww.tektronix.com077-0738-00

Strona 224 - REFLevel:ABSolute:HIGH

Command GroupsTable2-15:BusCommands(cont.)Commands DescriptionBUS:B<x>:POSition Sets or returns the position of the specifiedbus waveformBUS:B<

Strona 225

Commands Listed in Alphabetical OrderArgumentsTTL specifies a presetTTL high level of 1.4 V.ECL specifies a preset ECL high level of –1.3 V.SEARCH:SEARC

Strona 226 - REFLevel:ABSolute:MID2

Commands Listed in Alphabetical OrderSyntaxSELect {ON|OFF}ArgumentsON turns the selected waveform display on.OFF turns the selected waveform display o

Strona 227 - REFLevel:METHod

Commands Listed in Alphabetical OrderExamplesSELECT:CH2 ON turns the channel 2 waveform display on, and selects channel 2.SELECT:CH1? might return :SE

Strona 228

Commands Listed in Alphabetical OrderSyntaxSELect:D<x> {<NR1>|OFF|ON}SELect:D<x>?ArgumentsON turns on the display of the specified wa

Strona 229 - :REFLevel:PERCent:MID[1]

Commands Listed in Alphabetical OrderSELect:REF<x>Turns on and off the display of the reference waveform <x>. The <x > variablerepre

Strona 230

Commands Listed in Alphabetical OrderRelated CommandsHEADer, *LRN?,VERBoseExamplesSET? returns a long response, part of which could be as follows: :SE

Strona 231 - MESSage:BOX

Commands Listed in Alphabetical OrderSETUP<x>:LABELSets or returns the setup label for the specified channel <x>.GroupSave and RecallSyntax

Strona 232 - MESSage:SHOW

Commands Listed in Alphabetical OrderArguments<NR1> is a valuein the range from 0 through 255. The binary bits of the SRERare set according to t

Strona 233

Commands Listed in Alphabetical Order10000 : WFMINPRE:PT_FMT Y :WFMINPRE :XUNIT "s" :WFMINPRE:XINCR4.0000E-9 :WFMINPRE:XZERO 0.0000 :WFMINPR

Strona 234

Commands Listed in Alphabetical OrderSyntaxTIME <Q String>TIME?Related CommandsDATEArguments<QString> is a time in the form "hh:mm:ss

Strona 235 - NEWpass (No Query Form)

Command GroupsTable 2-15: Bus Commands (cont.)Commands DescriptionSEARCH:SEARCH<x>:TRIGger:A:BUS?Returns the serial search typeSEARCH:SEARCH<

Strona 236 - <QString> is th

Commands Listed in Alphabetical OrderTRIGger (No Query Form)Forces a trigger event to occur.GroupTriggerSyntaxTRIGger FORCeTRIGger?ArgumentsFORCe crea

Strona 237 - PICTBridge:DATEPrint

Commands Listed in Alphabetical Order800.0000E-3;:TRIGGER:A:LOWERTHRESHOLD:CH1 20.0000E-3;C H20.0000;CH3 0.0000;CH4 0.0000;:TRIGGER:A:HOLDOFF:TI ME20.

Strona 238 - PICTBridge:IDPrint

Commands Listed in Alphabetical OrderSyntaxTRIGger:A:BUS{I2C|SPI|CAN|RS232}TRIGger:A:BUS?ArgumentsI2C specifies the Inter-IC bus.SPI specifies the Seria

Strona 239 - MAGESize

Commands Listed in Alphabetical OrderTRIGger:A:BUS:B<x>:CAN:DATa:DIRectionSets or returns the CAN trigger data direction to be valid on a Read,

Strona 240 - APERSize

Commands Listed in Alphabetical OrderUNEQual sets the oscilloscope to trigger when the data is not equal to the qualifi ervalue.LESSEQual sets the osci

Strona 241 - PICTBridge:PAPERType

Commands Listed in Alphabetical OrderSyntaxTRIGger:A:BUS:B<x>:CAN:DATa:VALue <QString>TRIGger:A:BUS:B<x>:CAN:DAT a:VALue?Arguments&l

Strona 242 - RINTQual

Commands Listed in Alphabetical OrderConditionsThis command requires a DPO2AUTO application m odule.GroupTriggerSyntaxTRIGger:A:BUS:B<x>:CAN{:I

Strona 243 - *RCL (No Query Form)

Commands Listed in Alphabetical OrderSyntaxTRIGger:A:BUS:B<x>:I2C:ADDRess:MODe {ADDR7|ADDR10}TRIGger:A:BUS:B<x>:I2C:ADD Ress:MODe?Argument

Strona 244 - RECAll:SETUp (No Query Form)

Commands Listed in Alphabetical OrderSyntaxTRIGger:A:BUS:B<x>:I2C:ADDRess:VALue <QString>TRIGger:A:BUS:B<x>:I2C:AD DRess:VALue?Argum

Strona 245 - REF<x>? (Query Only)

Commands Listed in Alphabetical OrderTRIGger:A:BUS:B<x>:I2C:DATa:DIRectionSets or returns the I2C trigger type to be valid on a Read, Write, o r

Strona 246 - REF<x>:LABel

Command GroupsCalibration and Diagnostic Command GroupThe Calibration and Diagnostic commands provide information about the currentstate of o scillosc

Strona 247 - REF<x>:SCAle

Commands Listed in Alphabetical OrderTRIGger:A:BUS:B<x>:I2C:DATa:VALueSets or returns the binary data string used for I2C triggering if the trig

Strona 248 - REF<x>:VERTical:SCAle

Commands Listed in Alphabetical OrderExamplesTRIGGER:A:BUS:B1:LIN:CONDITION ERRORsets the LIN trigger conditionto error.TRIGGER:A:BUS:B1:LIN:CONDI TIO

Strona 249 - REM (No Query Form)

Commands Listed in Alphabetical OrderArgumentsLESSThan sets the LIN data qualifier to less than.MOREThan sets the LIN data qualifier to greater than.EQU

Strona 250 - *SAV (No Query Form)

Commands Listed in Alphabetical OrderConditionsThis command requires a DPO2AUTO application module.GroupBusSyntaxTRIGger:A:BUS:B<x>:LIN:DAT a:VA

Strona 251 - SAVe:ASSIgn:TYPe

Commands Listed in Alphabetical OrderTRIGger:A:BUS:B<x>:LIN:IDentifier:VALueSets or returns the binary address string used for LIN trigger if the

Strona 252 - SAVe:IMAGe:FILEFormat

Commands Listed in Alphabetical OrderGroupTriggerSyntaxTRIGger:A:BUS:B<x>:RS232C: CONDition{RXSTArt|RXDATA|RXENDPacke t|RXPARity|TXSTArt|TXDATA|

Strona 253 - SAVe:SETUp (No Query Form)

Commands Listed in Alphabetical OrderGroupTriggerSyntaxTRIGger:A:BUS:B<x>:RS232C :RX:DATa:VALueTRIGger:A:BUS:B<x>:RS232C :RX:DATa:VALue?Ar

Strona 254 - SAVe:WAVEfor

Commands Listed in Alphabetical OrderTRIGger:A:BUS:B<x>:SPI:CONDitionSets or returns the trigger condition for a SPI trigger. Applies t o bus &l

Strona 255 - SAVe:WAVEform:FILEFormat

Commands Listed in Alphabetical OrderTRIGger:A:BUS:B<x>:SPI:DATa{:OUT|:MOSI}:VALueSets or returns the binary data string to be used for a SPI tr

Strona 256

Commands Listed in Alphabetical OrderSyntaxTRIGger:A:BUS:SOUrce {B 1|B2}TRIGger:A:BUS:SOUrce?ArgumentsB1 specifies the Bus 1 source.B2 specifies the Bus

Strona 257 - SEARCH? (Query Only)

Command GroupsCursor CommandGroupUse the commands in the Cursor Command Group to control the cursor displayand readout. You can use these commands to

Strona 258

Commands Listed in Alphabetical OrderNOISErej specifies noise-rejection coupling, which provides stable triggeringby increa sing the trigger hysteresis

Strona 259

Commands Listed in Alphabetical OrderRelated CommandsTRIGger:A:EDGE:SLOpe, TRIGger:A:EDGE:COUPlingArgumentsCH1–CH4 specifies an analog input channel as

Strona 260 - SEARCH:SEARCH<x>:STATE

Commands Listed in Alphabetical OrderArguments<NR3> specifi esthe holdoff time in seconds. The range is from 20 ns through 8.0 s.ExamplesTRIGGER:

Strona 261

Commands Listed in Alphabetical OrderExamplesTRIGGER:A:LEVEL:AUXIN E CLsets the auxiliary input trigger level to -1.3volts.TRIGGER:A:LEVEL:AUXIN?might

Strona 262

Commands Listed in Alphabetical OrderArgumentsECL specifies a presetECL high level of –1.3 V.TTL specifies a preset TT L high level of 1.4 V.<NR3>

Strona 263 - SEARCH:S

Commands Listed in Alphabetical Order3,and4aremeton four-channel oscilloscopes. On two-channel oscilloscopes,only channel 1 and channel 2 are availabl

Strona 264

Commands Listed in Alphabetical OrderTRIGger:A:LOGIc:INPut? (Query Only)Returns the logic input values for all channels. If a clock channel is defined,

Strona 265

Commands Listed in Alphabetical OrderSyntaxTRIGger:A:LOGIc:INPut:CLOCk:EDGE {FALL|RISe|EITher}TRIGger:A:LOGIc:INPut:CLOC k:EDGE?ArgumentsRISe specifies

Strona 266 - SEARCH:SEARC

Commands Listed in Alphabetical OrderTRIGger:A:LOGIc:PATtern? (Query Only)Returns the conditions used for generating an A logic pattern trigger, with

Strona 267 - SEARCH:SEARCH<x>:TRI

Commands Listed in Alphabetical OrderGroupTriggerSyntaxTRIGger:A:LOGIc:PATtern:IN Put:D<x> {HIGH|LOW|X}TRIGger:A:LOGIc:PATtern:IN Put:D<x>

Strona 268

Command GroupsTable 2-17: Cursor Commands (cont.)CommandDescriptionCURSor:XY:POLar:RADIUS:POSITION<x>? Returns the polar radius of the specifiedc

Strona 269

Commands Listed in Alphabetical OrderTRIGger:A:LOGIc:PATtern:WHEn:LESSLimitSets or returns the maximum time that the selected pattern may be true and

Strona 270

Commands Listed in Alphabetical OrderTRIGger:A:LOGIc:THReshold:(CH<x>|D<x>)This command sets or queries the logic trigger threshold voltag

Strona 271

Commands Listed in Alphabetical OrderExamplesTRIGGER:A:LOWERTHRESHOLD:CH2 50E-3sets the lower limit threshold forCH2ofthepulserunttriggerto50mV.TRIGGE

Strona 272

Commands Listed in Alphabetical OrderTRIGGER:A:MODE?might return :TRIGGER :A:MODE NORMAL indicating that avalid trigger event must occur before a trig

Strona 273

Commands Listed in Alphabetical OrderExamplesTRIGGER:A:PULSE:CLASS WIDTHspecifies a width pulse for the A trigger.TRIGGER:A:PULSE:CLASS? might return :

Strona 274

Commands Listed in Alphabetical OrderGroupTriggerSyntaxTRIGger:A:PULSEWidth:SOUrc e {{CH1|CH2|CH3|CH4|D0|D1|D2|D3|D4|D5|D6|D7|D8 |D9|D10|D11|D12|D13|D

Strona 275

Commands Listed in Alphabetical OrderExamplesTRIGGER:A:PULSEWIDTH:WHEN LESST HANspecifies that the duration of the Apulse will fall within defined high

Strona 276

Commands Listed in Alphabetical OrderGroupTriggerSyntaxTRIGger:A:RUNT:POLarity {EITher|NEGati ve|POSitive}TRIGger:A:RUNT:POLarity?ArgumentsPOSitive in

Strona 277

Commands Listed in Alphabetical OrderGroupTriggerSyntaxTRIGger:A:RUNT:WHEn {LESSthan|MOREt han|EQual|UNEQu al|OCCURS}TRIGger:A:RUNT:WHEn?Related Comma

Strona 278

Commands Listed in Alphabetical OrderExamplesTRIGGER:A:RUNT:WIDTH 15E-6sets the minimum width of the pulse runttriggerto15µs.TRIGGER:A:RUNT:WIDTH?migh

Strona 279

Command GroupsTable 2-18: Display CommandsCommandDescriptionDISplay?Returns current display settingsDISplay:CLOCk Sets or returns the display of the d

Strona 280 - H<x>:TRIGger:A:LEVel

Commands Listed in Alphabetical OrderGroupTriggerSyntaxTRIGger:A:SETHold:CLOCk:E DGE {FALL|RISe}TRIGger:A:SETHold:CLOCk:E DGE?ArgumentsFALL specifies p

Strona 281

Commands Listed in Alphabetical OrderGroupTriggerSyntaxTRIGger:A:SETHold:CLOCk:TH Reshold {<NR3>|TTL|ECL}TRIGger:A:SETHold:CLOCk:TH Reshold?Argu

Strona 282

Commands Listed in Alphabetical OrderSyntaxDPO Models:TRIGger:A:SETHold:DATa:SO Urce {CH1|CH2|CH3|CH4|D0|D1|D2|D3|D4|D5|D6|D7|D 8|D9|D10|D11|D12|D13|D

Strona 283 - SEARCH:SE

Commands Listed in Alphabetical OrderTRIGger:A:SETHold:HOLDTimeSets or returns the hold time for setup and hold violation triggering.GroupTriggerSynta

Strona 284

Commands Listed in Alphabetical OrderGroupTriggerSyntaxTRIGger:A:SETHold:THResho ld:CH<x> {<NR3>|ECL|TTL}TRIGger:A:SETHold:THResho ld:CH&l

Strona 285

Commands Listed in Alphabetical OrderGroupTriggerSyntaxTRIGger:A{:TRANsition|:RIS EFall}?Related CommandsTRIGger:A:UPPerthreshold:CH<x> , TRIGge

Strona 286

Commands Listed in Alphabetical OrderArgumentsPOSitive indicates that a pulse edge must trave rse from the lower (mostnegative) to higher (most positi

Strona 287

Commands Listed in Alphabetical OrderSyntaxTRIGger:A{:TRANsition|:RISEFall}:WHEn{SLOWer|FASTer|EQual|UNEQu al}TRIGger:A{:TRANsition|:RIS EFall}:WHEn?A

Strona 288

Commands Listed in Alphabetical OrderLOGic specifi esthat a trigger occurs when specified conditions are met and iscontrolledbytheTRIGger:A:LOGIc? comma

Strona 289

Commands Listed in Alphabetical OrderGroupTriggerSyntaxTRIGger:A:VIDeo?ExamplesTRIGGER:A:VIDEO? might return :TRIGGER:A:VIDEO:POLARITYPOSITIVE;SOURCE

Strona 290

Command GroupsTable 2-19: Ethernet Commands (cont.)CommandDescriptionETHERnet:HTTPPortSets or returns the remote interface HTTPport valueETHERnet:IPAD

Strona 291

Commands Listed in Alphabetical OrderRelated CommandsTRIGger:A:VIDeo{:SYNC|:FIELD}Arguments<NR1> argument is an integer that sets the video line

Strona 292

Commands Listed in Alphabetical OrderGroupTriggerSyntaxTRIGger:A:VIDeo:SOUrce {CH1|CH2|CH3|CH 4}TRIGger:A:VIDeo:SOUrce?ArgumentsCH1–CH4 specifies the i

Strona 293

Commands Listed in Alphabetical OrderGroupTriggerSyntaxTRIGger:A:VIDeo{:SYNC|:FI ELD}{ODD|EVEN|ALLFields|ALLLi nes|NUMERic}TRIGger:A:VIDeo{:SYNC|:FI E

Strona 294

Commands Listed in Alphabetical OrderExamplesTRIGGER:EXTERNAL:PROBE?might return :TRIGGER:EXTERNAL:PROBE1.0E1for a 10X probe.TRIGger:EXTernal:YUNIts?

Strona 295

Commands Listed in Alphabetical OrderAUTO indicatesthat the oscilloscope is in the automatic mode and acquires dataeven in the abs ence of a trigger.R

Strona 296

Commands Listed in Alphabetical OrderUSBDevice:CONFigureEnables or disables the rear USB port for use with Pictbridge printers.GroupPictBridgeSyntaxUS

Strona 297

Commands Listed in Alphabetical OrderUSBTMC:PRODUCTID:HEXadecimal? (Query Only)Returns the product ID of the USBTMC device in hexadecimal format.Group

Strona 298

Commands Listed in Alphabetical OrderGroupMiscSyntaxUSBTMC:VENDORID:HEXadecima l?ExamplesUSBTMC:VENDORID:HEXADECIMA L? might returnUSBTMC:VENDORID:HEX

Strona 299

Commands Listed in Alphabetical Order*WAI (No QueryForm)Prevents the o scilloscope from executing further commands or queries until allpending command

Strona 300

Commands Listed in Alphabetical Order“V”;YMULT 4.0000E-3;YOFF 0.0000;YZERO 0.0000;:CURVE2,1,4,2,4,3,0,3,3,3,3,3,3, 4,3,5,6,6,7,3WFMInpre? (Query Only)

Strona 301 - SELect:CH<x>

Command GroupsNOTE. Using back slas h as a path separator may produce some unexpectedresults, depending on how your application treats escaped charact

Strona 302 - SELect:D<x>

Commands Listed in Alphabetical OrderWFMInpre:BN_FmtSets or returns the format of binary data for incoming waveforms.GroupWaveform TransferSyntaxWFMIn

Strona 303 - SELect:MATH[1]

Commands Listed in Alphabetical OrderWFMInpre:BYT_OrSets or returns which byte of binary waveform data is expected first for incomingwaveform data when

Strona 304 - SET? (Qu

Commands Listed in Alphabetical OrderExamplesWFMINPRE:COMPOSITION?might return :WFMINPRE:COMPOSI TIONSINGULAR_YTindicating that the sample part of sin

Strona 305

Commands Listed in Alphabetical OrderExamplesWFMINPRE:FILTERFREQ?might return :WFMINPRE:FILTERF REQ 200000000indicating that the FilterVu frequency fo

Strona 306 - SETUP<x>:LAB

Commands Listed in Alphabetical OrderArgumentsENV spec ifies that the waveform is transmitted in envelope mode as maximumand minimum point pairs. OnlyY

Strona 307 - TEKSecure (No Query Form)

Commands Listed in Alphabetical OrderSyntaxWFMInpre:XINcr <NR3>WFMInpre:XINcr?Related CommandsWFMInpre:XUNit, WFMOutpre:XINcr?Arguments<NR3&g

Strona 308

Commands Listed in Alphabetical OrderWFMInpre:XZEroSets or returns the position value, in XUNits, of the first sample of the incomingwaveform.GroupWave

Strona 309 - *TRG (No Query Form)

Commands Listed in Alphabetical OrderRelated CommandsDATa:DESTination, WFMInpre:BYT_Nr, WFMInpre:YUNitArguments<NR3> is the vertical scale facto

Strona 310 - TRIGger:A

Commands Listed in Alphabetical OrderWFMINPRE:YOFF?might return :WFMINPRE: YOFF 25 indicating that the verticalposition of the incoming waveform in di

Strona 311 - TRIGger:A:BUS

Commands Listed in Alphabetical OrderGroupWaveform TransferSyntaxWFMInpre:YZEro <NR3>WFMInpre:YZEro?Related CommandsWFMInpre:YUNit, WFMOutpre:YZ

Strona 312

Command GroupsTable 2-21: FilterVuCommands (cont.)CommandDescriptionFILTERVu:FREQuency Sets or queries the FilterVu frequencyFILTERVu:FREQuency:AVAILa

Strona 313 - TRIGger:A:BU

Commands Listed in Alphabetical Order4.0000E-9;XZERO - 20.0000E-6;PT_OFF 0;YUNIT “V”;YMULT15.6250E-6;YOFF :”6.4000E+3;YZERO 0.0000WFMOutpre:BIT_NrSets

Strona 314

Commands Listed in Alphabetical OrderArgumentsRI specifies signed integer data point representation.RP specifies positive integer data point representat

Strona 315

Commands Listed in Alphabetical OrderSyntaxWFMOutpre:BYT_Or {LSB| MSB}WFMOutpre:BYT_Or?Related CommandsWFMOutpre:ENCdg, WFMOutpre:BYT_NrArgumentsLSB s

Strona 316

Commands Listed in Alphabetical OrderSyntaxWFMOutpre:ENCdg {ASCii| BINary}WFMOutpre:ENCdg?Related CommandsDATa:ENCdg, WFMOutpre:BYT_Nr, WFMInpre:FILTE

Strona 317

Commands Listed in Alphabetical OrderGroupWaveform TransferSyntaxWFMOutpre:NR_Pt?Related CommandsCURVe, DATa, DATa:STARt, DATa:STOP, SAVe:WAVEform,SAV

Strona 318

Commands Listed in Alphabetical OrderRelated CommandsDATa:SOUrceArgumentsArguments are ignored.ExamplesWFMOUTPRE:PT_OFF? might return WFMOUTPRE:PT_OFF

Strona 319

Commands Listed in Alphabetical Ordertime out and an error will be generated if the waveform specified by DATa:SOUrceis not turned on.GroupWaveform Tra

Strona 320 - S:B<x>:I2C:DATa:VALue

Commands Listed in Alphabetical OrderWFMOutpre:XINcr? (Query O nly)Returns the horizontal point spacing in units of WFMOutpre:XUNit for thewaveform sp

Strona 321

Commands Listed in Alphabetical OrderGroupWaveform TransferSyntaxWFMOutpre:XZEro?Related CommandsDATa:SOUrce, WFMOutpre:XUNit?ExamplesWFMOUTPRE:XZERO?

Strona 322

Commands Listed in Alphabetical OrderSyntaxWFMOutpre:YOFf?Related CommandsDATa:SOUrce, WFMOutpre:BYT_NrExamplesWFMOUTPRE:YOFF? might return :WFMOUTP R

Strona 323

Command GroupsHorizontal Command GroupUse the commands in the Horizontal Command Group to control the oscilloscopetime bases. You can set the time-per

Strona 324

Commands Listed in Alphabetical OrderExamplesWFMOUTPRE:YZERO?might return :WFMOUTPRE:YZERO -100.0000E-3indicating that vertical offset is set to –100

Strona 325

Commands Listed in Alphabetical OrderZOOm:ZOOM1? (Query Only)Returns the current horizontal positioning and scaling of the display.GroupZoomSyntaxZOOm

Strona 326

Commands Listed in Alphabetical OrderZOOM:ZOOM1:HORIZONTAL:POSITION?might return:ZOOM1:HORIZONTAL:POSITIO N 50.0000 indicating that the Zoom1reference

Strona 327 - S:B<x>:SPI:CONDition

Commands Listed in Alphabetical OrderSyntaxZOOm:ZOOM1:SCAle <NR3>ZOOm:ZOOM1:SCAle?Arguments<NR3> is the amount of expansion in the horizon

Strona 328 - TRIGger:A:BUS:SOUrce

Commands Listed in Alphabetical Order2-378 MSO2000B, DPO2000B, M SO2000 and DPO2000 Series Oscilloscopes Programmer Manual

Strona 329 - TRIGger:A:EDGE:COUPling

Status and EventsThe oscilloscope provides a status and event reporting system for the Ethernet,GPIB (with the TEK-USB-488 Adapter), and USB interfac

Strona 330 - TRIGger:A:EDGE:SOUrce

Status and EventsTable 3-1: SESRBit FunctionsBit F unction7 (MSB) PON Power On. Shows that the oscilloscope was powered on.On completion, the diagnost

Strona 331 - TRIGger:A:HOLDoff:TIMe

Status and EventsTable 3-2: SBR Bit Functions (cont.)Bit Function4MAVMessage Available. Shows that output is availablein the Output Queue.3————Not use

Strona 332 - TRIGger:A:LEVel:AUXin

Status and EventsUse the *SRE command to set the SRER. Use the *SRE? query to read theregister. The RQS bit remains set to one until either the Status

Strona 333 - TRIGger:A:LEVel:D<x>

Status and EventsRead the Event Queue with the EVENT? query (which returns only the eventnumber), with the EVMSG? query (which returns the event numbe

Strona 334 - TRIGger:A:LOGIc:CLAss

Copyright © Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiariesor suppliers, and are protected by na

Strona 335 - TRIGger:A:LOGIc:FUNCtion

Command GroupsTable 2-24: Mark CommandsCommand DescriptionMARK Move to the next or p revious mark on thewaveform or returns a ll learnable settingsfro

Strona 336 - GIc:INPut? (Query Only)

Status and EventsFigure 3-6: Status and Event Handling ProcessWhen an event occurs, a signal is sent to the DESER (1). If that type of eventis enabled

Strona 337

Status and EventsSynchronization MethodsOverviewAlthough most commands are completed almost immediately after being receivedby the oscilloscope, some

Strona 338 - GIc:PATtern? (Query Only)

Status and EventsFor example, a typical application m ight involve acquiring a single-sequencewaveform and then taking a measurement on the acquired w

Strona 339 - TRIGger:A:LOGIc:PATtern:W HEn

Status and EventsThesamecommand sequence using the *WAI command for synchronizationlooks like this:/* Set up conditional ac quisition */ACQUIRE:STATE

Strona 340 - GIc:PATtern:WHEn:LESSLimit

Status and Eventsmany commands to the input buffer. The controller is still tied up though, andthe repeated BUSY? query will result in bus traffic.Usin

Strona 341

Status and EventsThesamecommand sequence using the *OPC command for synchronizationlooks like this/* Set up conditional acquisition */ACQUIRE:STATE OF

Strona 342 - TRIGger:A:MODe

Status and Events/* Wait until the acquis ition is complete */*OPC?/* Take amplitude measurement */MEASUREMENT:IMMED:VALUE?This is the simplest approa

Strona 343 - TRIGger:A:PULse:CLAss

Status and EventsTable 3-4: Command Error Messages (CME Bit 5) (cont.)Code Message108 Parameter not allowed109 Missing parameter110Command header erro

Strona 344 - TRIGger:A:PULSEWidth:SOUrce

Status and EventsTable 3-5: Execution Error Messages (EXE B it 4) (cont.)Code Message253Corrupt media254Media full255Directory full256File name not fo

Strona 345 - TRIGger:A:PULSEWidth:WHEn

Status and EventsTable 3-5: Execution Error Messages (EXE Bit 4) (cont.)Code Message2220 Measurement error, No valid edge - No arm cross2221 Measureme

Strona 346 - TRIGger:A:RUNT:POLarity

Command GroupsMath Command GroupUse the commands in the Math Command Group to create and define a mathwaveform. Use the available math functions to defi

Strona 347 - TRIGger:A:RUNT:WHEn

Status and EventsDevice ErrorThe following table lists the device errors that can occur d uring oscilloscopeoperation. These errors may indicate that

Strona 348 - TRIGger:A:RUNT:WIDth

Status and EventsTable 3-8: Execution Warning Messages (EXE Bit 4) (cont.)Code Message543 Measurement warning, Low resolution544 Measurement warning,

Strona 349 - TRIGger:A:SETHold:CLOCk:EDGE

Status and Events3-18 MSO2000B, DPO2000B, MSO2000 and DPO2000 Series Oscilloscopes Programmer Manual

Strona 350

Appendix A: Character SetMSO2000B, DPO2000B, MSO2000 and DPO2000 Series Oscilloscopes Programmer Manual A-1

Strona 351 - TRIGger:A:SETHold:DATa:SOUrce

Appendix A: Character SetA-2 MSO2000B, DPO2000B, MSO2000 and DPO2000 Series Oscilloscopes Programmer M anual

Strona 352

Appendix B: Waveform Data in MSO/DPO2000B andMSO/DPO2000 Series InstrumentsAn analog channel of an MSO/DPO2000B and MSO/DPO2000 seriesinstrument has a

Strona 353 - TRIGger:A:SETHold:SETTime

Appendix B: Waveform Data in MSO/DPO2000B and MSO/DPO2000 Series InstrumentsComposition of WaveformDataThe MSO/DPO2000/B series introduces the concept

Strona 354

Appendix B: Wa veform Data in MSO/DPO2000B and MSO/DPO2000 Series InstrumentsCOMPOSITE_YTCOMPOSITE_YT, COMPOSITE_ENVSINGULAR_YTThe composition of the

Strona 355

Appendix B: Waveform Data in MSO/DPO2000B and MSO/DPO2000 Series Instrumentsreduced resolution waveform. The filtering frequency of the full resolution

Strona 356

Appendix B: Wa veform Data in MSO/DPO2000B and MSO/DPO2000 Series InstrumentsQuerying AcquisitionPeak Detect Data from theInstrument1. Use the DATa:CO

Strona 357 - TRIGger:A:TYPe

Command GroupsTable 2-25: Math Commands (cont.)CommandDescriptionMATH[1]:SPECTral:NYQUISTFreq? Returns the Nyquist frequency of the FFTmath waveformMA

Strona 358 - TRIGger:A:VIDeo? (Query Only)

Appendix B: Waveform Data in MSO/DPO2000B and MSO/DPO2000 Series Instrumentsturned on. Use the WFMOutpre:RECOrdlength? query to determine the a ctualr

Strona 359 - TRIGger:A:VIDeo:LINE

Appendix C: Reserved WordsThis is a list of reserved words for your instrument. Capital letters identify therequired minimum spelling.*CAL*CLS*DDT*ESE

Strona 360 - TRIGger:A:VIDeo:SOUrce

Appendix C: Reserved WordsDATABitsDATEDATEPrintDATaDAllDBDCDECimalDEFLTDEFaultDEFineDEGAUssDEGreesDELEteDELIMiterDELTaDELTatimeDELayDELayedDELeteDELta

Strona 361 - TRIGger:A:VIDeo{:SYNC

Appendix C: Reserved WordsLETTERLEVelLFLFRejLINLINELINEArLINEPeriodLISTLOCkLOGLOGICLOGIcLOGicLOOPLOWLOWLimitLOWerthresholdLSBMAGMAINMAInMAIn]MARKMARKS

Strona 362 - TRIGger:EXTernal:PRObe

Appendix C: Reserved WordsSAMPLERateSAMPLEpointSAMpleSAVeSCANSCAleSCLKSCREENSCREenSDATASEARCHSEARCHtotriggerSECAMSECondsSELectSELectedSEQuenceSERIALnu

Strona 363 - TRIGger:STATE? (Query Only)

Appendix D: Application Module-enabled CommandsThis table lists the c ommands that become available with installation of anapplication module.MSO2000B

Strona 364 - UNLock (No Query F orm)

Appendix D: Application Module-enabled CommandsApplication ModuleCommandsAutomotive Serial Triggering and Analysis Module. Enables triggering on packe

Strona 365 - USBDevice:CO

Appendix D: Application Module-enabled CommandsApplication M oduleCommandsDPO2COMP Computer Serial Triggering and Analysis Module. Enables triggering

Strona 366 - USBTMC:PRODU

Appendix D: Application Module-enabled CommandsApplication ModuleCommandsDPO2EMBD Embedded Serial Triggering and Analysis Module. Enables triggering o

Strona 367

Appendix E: Search and Trigger Command SequenceExamplesThe following are some example command sequences that show a variety ofdifferent sear ches and

Strona 368 - WAVFrm? (Query Only)

Command GroupsTable 2-26: Measurement Commands (cont.)CommandDescriptionMEASUrement:IMMed:DELay? Returns information about the immediatedelay measurem

Strona 369 - WFMInpre:BIT_Nr

Appendix E: Search and Trigger Command Sequence ExamplesCommand Comment:TRIGger:A:TYPe EDGE Specifies that this will be an edge trigger (trigger will o

Strona 370 - WFMInpre:BYT_Nr

Appendix E: Search and Trigger Command Sequence ExamplesCommand Comment*RST Resets the oscilloscope. Wait for the reset to complete(approximately 3 se

Strona 371 - WFMInpre:COMPosition

Appendix E: Search and Trigger Command Sequence ExamplesE-4 MSO2000B, DPO2000B, MSO2000 and DPO2000 Series Oscilloscopes Programmer Manual

Strona 372 - WFMInpre:FILTERFreq

IndexAACQuire:MAXSamplerate?, 2-59ACQuire:MODe, 2-59ACQuire:NUMACq?, 2-60ACQuire:NUMAVg, 2-61ACQuire:STATE, 2-61ACQuire:STOPAfter, 2-63ACQuire?, 2-59A

Strona 373 - WFMInpre:PT_Fmt

IndexC*CAL?, 2-95CALibrate:FACtory, 2-95CALibrate:FACtory:STATus?, 2-96CALibrate:INTERNal, 2-96CALibrate:INTERNal:STARt, 2-96CALibrate:INTERNal:STATu

Strona 374 - WFMInpre:XINcr

Index*DDT, 2-134DESE, 2-135DIAg:LOOP:OPTion, 2-135DIAg:LOOP:OPTion:NTIMes, 2-136DIAg:LOOP:STOP, 2-136DIAg:RESUlt:FLAg?, 2-137DIAg:RESUlt:LOG?, 2-137DI

Strona 375 - WFMInpre:XUNit

IndexLLANGuage, 2-172LOCk, 2-172*LRN?, 2-173MMark Command Group, 2-23MARK, 2-174MARK:CREATE, 2-174MARK:DELEte, 2-175MARK:FREE?, 2-175MARK:SELected:END

Strona 376 - WFMInpre:YMUlt

IndexMiscellaneousCommand Group, 2-29NNEWpass, 2-219O*OPC, 2-220PPASSWord, 2-221PICTBridge:DATEPrint, 2-221PICTBridge:DEFault, 2-222PICTBridge:IDPrint

Strona 377 - WFMInpre:YOFf

IndexSEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:LIN:DATa:VALue, 2-257SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:LIN:ERRTYPE, 2-257SEARCH:SEA

Strona 378 - WFMInpre:YZEro

IndexSEARCH:SEARCH<x>:TRIGger:A:RUNT:WHEn, 2-276SEARCH:SEARCH<x>:TRIGger:A:RUNT:WIDth, 2-277SEARCH:SEARCH<x>:TRIGger:A:SETHold:CLOCk

Strona 379 - WFMOutpre? (Query Only)

Command GroupsTable 2-26: Measurement Commands (cont.)CommandDescriptionMEASUrement:MEAS<x>:MEAN?Returns the mean value accumulated sinc ethe la

Strona 380 - WFMOutpre:BN_Fmt

IndexTRIGger:A:BUS:B<x>:RS232C:TX:DATa:SIZe, 2-310TRIGger:A:BUS:B<x>:RS232C:TX:DATa:VALue, 2-310TRIGger:A:BUS:B<x>:SPI:CONDition, 2-

Strona 381 - WFMOutpre:BYT_Or

IndexUSBTMC:SERIALnumber?, 2-350USBTMC:VENDORID:DECimal?, 2-350USBTMC:VENDORID:HEXadecimal?, 2-350VVERBose, 2-351Vertical Command Group, 2-47W*WAI, 2-

Strona 382 - WFMOutpre:ENCdg

Command GroupsTable 2-26: Measurement Commands (cont.)CommandDescriptionMEASUrement:MEAS<x>:SOUrce<x> Sets or returns the source for the s

Strona 383 - WFMOutpre:NR_Pt? (Query Only)

Command GroupsTable 2-27: Miscellaneous Commands (cont.)CommandDescriptionLOCk Sets or returns the front panel lock state*LRN? Returns a listing of o

Strona 384 - WFMOutpre:PT_Off?

Command GroupsTable 2-28: PictBridge Commands (cont.)CommandDescriptionPICTBridge:IDPrintEnables or disables printing the o scilloscopemodel and seria

Strona 385 - WFMOutpre:WFId? (Query Only)

Command GroupsTable 2-29: Save and Recall Commands (cont.)CommandDescriptionSAVe:IMAGe:LAYout Sets or returns the layout to use for savedscreen images

Strona 386

Command GroupsSearch CommandGroupUse the commands in the Search Commands Group to seek out information inwaveform records.Search CommandsCommand Descr

Strona 387 - WFMOutpre:XZEro? (Query Only)

Table of ContentsGetting Started ... ... ... ... ... 1-1Setting U

Strona 388 - WFMOutpre:YOFf? (Query Only)

Command GroupsSearch Commands (cont.)Command DescriptionSEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:I2C:ADDRess:MODeSets or returns the I2C addres

Strona 389 - WFMOutpre:YZEro? (Query Only)

Command GroupsSearch Commands (cont.)Command DescriptionSEARCH:SEARCH<x>:TRIGger:A:EDG E:SOUrceSets or returns the source waveform for anedge se

Strona 390 - ZOOm{:MODe

Command GroupsSearch Commands (cont.)Command DescriptionSEARCH:SEARCH<x>:TRIGger:A:LOGIc:THReshold:REF<x>Sets or returns the reference w a

Strona 391 - Rizontal:POSition

Command GroupsSearch Commands (cont.)Command DescriptionSEARCH:SEARCH<x>:TRIGger:A:SETHold:THReshold:REF<x>Sets or returns the trigger sea

Strona 392 - ZOOm:ZOOM1:SCAle

Command GroupsTable 2-30: Status and Error Commands (cont.)CommandDescription*ESE Sets or returns the bits in the Event StatusEnable Register*ESR? Ret

Strona 393 - ZOOm:ZOOM1

Command Groupsset the threshold voltage levels using TRIGger:A:LEVel:AUXin,TRIGger:A:LEVel:CH<x>,orTRIGger:A:LEVel:D<x>.Logic TriggerLogic

Strona 394

Command GroupsRunt triggering lets you trigger on a pulse amplitude that crosses one thresholdbut fails to cross a second threshold before recrossing

Strona 395 - Status and Events

Command GroupsTable 2-31: Trigger CommandsCommandDescriptionTRIGgerForces a trigger event tooccurTRIGger:FREQuency?Returns the available triggerfreque

Strona 396

Command GroupsTable 2-31: Trigger Commands (cont.)CommandDescriptionTRIGger:A:BUS :B<x>:LIN:ID entifier:VALue Sets or returns the binaryaddress s

Strona 397 - Enable Registers

Command GroupsTable 2-31: Trigger Commands (cont.)CommandDescriptionTRIGger:A:BUS:B <x >:SPI:D ATa{:OUT|:M OSI}:VALue Sets or returns the binary

Strona 398 - Event Queue

Table of ContentsCommands Listed in Alphabetical Order ... 2-59Status and Events ...

Strona 399 - Event Handling Sequence

Command GroupsTable 2-31: Trigger Commands (cont.)CommandDescriptionTRIGger:A:LOGIc:INPut:C H<x> Specifies or returns thelogic setting for the sp

Strona 400

Command GroupsTable 2-31: Trigger Commands (cont.)CommandDescriptionTRIGger:A:PULse:CLAss Sets or returns the type ofpulse on which to triggerTRIGger:

Strona 401 - Synchronizat

Command GroupsTable 2-31: Trigger Commands (cont.)CommandDescriptionTRIGger:A:SETHold:DATa?Returns the voltagethreshold and data sourcefor the setup/h

Strona 402 - Using the *WAI Command

Command GroupsTable 2-31: Trigger Commands (cont.)CommandDescriptionTRIGger:A:VIDeo:POLarity Sets or returns the polarity ofthe A video triggerTRIGger

Strona 403 - Using the BUSY Query

Command GroupsTable 2-32: Ve rtical Commands (cont.)Command DescriptionAUXin:PRObe:ID:SERnumber? Returns the serial number of the probe thatis attache

Strona 404 - Using the *OPC Command

Command GroupsTable 2-32: Ve rtical Commands (cont.)Command DescriptionCH<x>:PRObe:ID:SERnumber? Returns the serial number of the probe thatis a

Strona 405

Command GroupsTable 2-32: Ve rtical Commands (cont.)Command DescriptionSELect:BUS<x> Turns on or off the specified bus waveformor returns whether

Strona 406 - Messages

Command GroupsNOTE. The source w aveform (as specified by the DATa:SOUrce) must be turnedon for the following queries to return a response:CURVe?DATa:C

Strona 407 - Execution Error

Command GroupsEach data valuerequires two to seven characters. This includes one to fivecharacters to represent the value, another charac t er, if the

Strona 408

Command Groupscommand. Next,specify the fi rst data point within the waveform record. Forexample, when you set DATa:STARt to 1, data points will be sto

Strona 409

Getting StartedThis manual explains the use of commands for remotely controlling youroscilloscope. With this information, you can write computer progr

Strona 410 - Execution Warning

Command GroupsTransferring WaveformData to the OscilloscopeYou can transfer waveforms to the oscilloscope from an external controller usingthe followi

Strona 411 - Internal Warning

Command GroupsTable 2-34: Waveform Transfer Commands (cont.)CommandDescriptionDATa:STOP Sets or returns the ending data point inwaveform transferWA VF

Strona 412

Command GroupsTable 2-34: Waveform Transfer Commands (cont.)CommandDescriptionWFMInpre:YZEroSets or returns the vertical offset of theincoming wavefor

Strona 413 - Appendix A: Character Set

Command GroupsTable 2-34: Waveform Transfer Commands (cont.)CommandDescriptionWFMOutpre:YMUlt? Returns the vertical scale factor per digitizinglevel f

Strona 414

Command Groups2-58 MSO2000B, DPO2000B, MSO2000 and DPO2000 Series Oscilloscopes Programmer Manual

Strona 415 - Record Length

Commands Listed in Alphabetical OrderACQuire? (Query Only)Returns the following current acquisition parameters:Stop afterAcquisition stateModeNumber o

Strona 416 - Composition of Waveform

Commands Listed in Alphabetical OrderWaveforms are the displayed d ata point values taken from acquisition intervals.Each acquisition interval represe

Strona 417 - Filtering of Waveform Data

Commands Listed in Alphabetical Orderstart acquisitions again, it will continue from 5000. The number will reset to 0only if you change the horizontal

Strona 418 - Displayed Waveform

Commands Listed in Alphabetical Ordersingle sequence acquisition will be started. If the last acquisition was continuous,a new continuous acquisition

Strona 419

Commands Listed in Alphabetical OrderACQuire:STOPAfterSets or returns whether the oscilloscope continually acquires acquisitions oracquires a single s

Strona 420

Getting StartedDPO2022B – 2 00 MHz, 2 channelSetting Up Remote Communications HardwareYou can remotely communicate between your oscilloscope and PC vi

Strona 421 - Appendix C: Reserved Words

Commands Listed in Alphabetical OrderArgumentsOFF turns aliasexpansion off. If a defined alias is sent when ALIas is off, acommand error (110) will be

Strona 422

Commands Listed in Alphabetical OrderRelated CommandsALIas[:STATE]ArgumentsThe first <QString> is the alias label.This label cannot be a command

Strona 423

Commands Listed in Alphabetical OrderGroupAliasSyntaxALIas:DELEte:ALLRelated CommandsALIas:DELEte, ALIas:DELEte[:NAMe]ExamplesALIAS:DELETE:ALL deletes

Strona 424

Commands Listed in Alphabetical OrderExamplesALIAS[:STATE]OFFturns the command alias feature off.ALIAS[:STATE]? returns 0 when the alias feature is of

Strona 425

Commands Listed in Alphabetical OrderAUTOSet:ENAbleEnables or disables the autoset.GroupMiscellaneousSyntaxAUTOSet:ENAble {OFF|ON}AUTOSet:ENAble?Relat

Strona 426

Commands Listed in Alphabetical OrderExamplesAUXIN:PROBE? might returnAUXIN:PROBE:ID:TYPE "No ProbeDetected";SERNUMBER "";:AUXIN:P

Strona 427

Commands Listed in Alphabetical OrderAUXin:PRObe:DEGAUss (No Query Form)Starts a degauss/autozero cycle on a TekVPI current probe attached to the A ux

Strona 428

Commands Listed in Alphabetical OrderArguments<NR3> is the probe range, which is probe dependent.AUXin:PRObe:GAINSets or returns the gain factor

Strona 429

Commands Listed in Alphabetical OrderGroupVerticalSyntaxAUXin:PRObe:RESistance?ExamplesAUXin:PRObe:RESistance? might return :AUXin:PR OBE:RESISTANCE1.

Strona 430

Commands Listed in Alphabetical OrderBUSSets or returns the parameters for each bus. These parameters affect either theSerial Trigger Setup or the Bus

Strona 431

Getting StartedTo view or change the Ethernet settings on your oscilloscope, do the following:1. On the front panel, push Utility.2. Push Utility Page

Strona 432

Commands Listed in Alphabetical OrderSyntaxBUS:B<x>:CAN:PRObe {CA NH|CANL|RX|TX|DI FFerential}BUS:B<x>:CAN:PRObe?ArgumentsCANH specifies th

Strona 433

Commands Listed in Alphabetical OrderArgumentsCH1–CH4 is the analog c hannel to use as the data source.D0–D15 is the digital channel to use as the dat

Strona 434

Commands Listed in Alphabetical OrderOFF does not include the read/write bit in the address.ON includes the read/write bit in the address.ExamplesBUS:

Strona 435

Commands Listed in Alphabetical OrderArgumentsCH1–CH4 specifies the analog channel to use as the I2C SDATA source.D0–D15 specifies the digital channel t

Strona 436

Commands Listed in Alphabetical OrderConditionsThis command requires a DPO2AUTO application m odule.GroupBusSyntaxBUS:B<x>:LIN:IDFORmat {NOPARit

Strona 437

Commands Listed in Alphabetical OrderConditionsThis command requires a DPO2AUTO application module.GroupBusSyntaxBUS:B<x>:LIN:SAMPLEpoint <NR

Strona 438

Commands Listed in Alphabetical OrderGroupBusSyntaxBUS:B<x>:LIN:STANDard {V1X|V2X|MIXed}BUS:B<x>:LIN:STANDard?ArgumentsV1X sets the LIN st

Strona 439

Commands Listed in Alphabetical OrderArgumentsEIther specifies either rising or falling edge as the clock edge.RISing specifies the rising edge as the c

Strona 440

Commands Listed in Alphabetical OrderSyntaxBUS:B<x>:PARallel:WIDth <NR1>BUS:B<x>:PARallel:WIDth?Arguments<NR1> is the number o

Strona 441

Commands Listed in Alphabetical OrderBUS:B<x>:RS232C:DATABitsSets or returns the number of RS232 data bits for bus <x>, where x is the bus